ERT for Medical Device According DIN EN 60601-1
General Technical Data
Protection degree | IP00 |
Isolation class | B |
Ambient temperature | max. +45°C |
Test voltage shaft - winding | 4 kV |
Frequenzy | 50/60 Hz |
Nominal Input voltage | 3x400V |
Altitude | 1000m |
Aprobation | CE,RoHs |
Design according | DIN EN 60601-1 |
Install | Ready for medical devices |
Series Medical Isolating Transformers Type G
Type | ERT 230/230/1G | ERT 230/230/2G | ERT 115/115/4G | ERT 230/230/4G | ERT 230/230/6G | ERT 230/230/10G |
---|---|---|---|---|---|---|
Input | 230V 50-60 Hz IEC-Socket |
230V 50-60 Hz IEC-Socket |
115V 50-60 Hz switchable 230V IEC-Socket |
230V 50-60 Hz IEC-Socket |
230V 50-60 Hz IEC-Socket |
230V 50-60 Hz IEC-Socket |
Input current A | 1,2 | 2,35 | 4,8 | 4,8 | 6,9 | 11 |
Output | 230V IEC-Socket 4x |
230V IEC-Socket 4x |
115V switchable to 230V IEC-Socket 4x |
230V IEC-Socket 4x |
230V IEC-Socket 8x |
230V IEC-Socket 8x |
Output Current A | 1 | 2 | 4 / 2 | 4 | 6 | 10 |
Fuse PRI | G 5x20 1,5A/T |
G 5x20 2,5A/TT |
G 5x20 4A/TT |
G 5x20 4A/TT |
G 5x20 8A/T |
G 6,3x32 12,5A/T |
Fuse SEC | G 5x20 1A/T |
G 5x20 2A/T |
G 5x20 4A/T |
G 5x20 4A/T |
G 5x20 6,3A/T |
G 5x20 10A/T |
Protection class | I | I | I | I | I | I |
Protection degree | IP20 | IP20 | IP20 | IP20 | IP20 | IP20 |
Power switch | 2-poles with Inrush limiter |
2-poles with Inrush limiter |
2-poles with Inrush limiter |
2-poles with Inrush limiter |
2-poles with Inrush limiter |
2-poles with Inrush limiter |
Energize mA | <0,3 | <0,3 | <0,3 | <0,5 | <0,5 | <0,5 |
Test voltage kV | 4 | 4 | 4 | 4 | 4 | 4 |
PRI-Shield kV | 2 | 2 | 2 | 2 | 2 | 2 |
SEC-Shield kV | 2 | 2 | 2 | 2 | 2 | 2 |
Ambient temperature | +40°C | +40°C | +40°C | +40°C | +40°C | +40°C |
Housing | Steel sheet | Steel sheet | Steel sheet | Steel sheet | Steel sheet | Steel sheet |
Dimensions (BxHxT) mm | 165x110x220 | 165x110x220 | 200x125x310 | 250x165x310 | 250x165x370 | 250x165x370 |
Mass kg | 6 | 6,6 | 11,7 | 10 | 17 | 26,7 |
Abrobation | CE, RoHS | CE, RoHS | CE, RoHS | CE, RoHS | CE, RoHS | CE, RoHS |
Series Medical Isolating Transformers Type W with ISO-Monitor
Type | ERT 230/230/1W | ERT 230/230/2W | ERT 230/230/4W | ERT 230/230/6G | ERT 230/230/10G |
---|---|---|---|---|---|
Isolation Monitor Pickup Signalisation, automatic test modus, keys for acknowledge and test |
≤53kΩ (±5) acoustically and optically available available |
≤53kΩ (±5) acoustically and optically available available |
≤53kΩ (±5) acoustically and optically available available |
≤53kΩ (±5) acoustically and optically available available/6G |
≤53kΩ (±5) acoustically and optically available available |
Input | 230V 50-60 Hz IEC-Socket |
230V 50-60 Hz IEC-Socket |
230V 50-60 Hz IEC-Socket |
230V 50-60 Hz IEC-Socket |
230V 50-60 Hz IEC-Socket |
Input current A | 1,2 | 2,35 | 4,8 | 6,9 | 11 |
Output | 230V IEC-Socket 4x |
230V IEC-Socket 4x |
230V IEC-Socket 4x |
230V IEC-Socket 8x |
230V IEC-Socket 8x |
Output current A | 1 | 2 | 4 | 6 | 10 |
Fuse PRI | G 5x20 1,5A/T |
G 5x20 2,5A/TT |
G 5x20 4A/TT |
G 5x20 8A/T |
G 6,3x32 12,5A/T |
Fuse SEC | G 5x20 1A/T |
G 5x20 2A/T |
G 5x20 4A/T |
G 5x20 6,3A/T |
G 5x20 10A/T |
Protection class | I | I | I | I | I |
Protection degree | IP20 | IP20 | IP20 | IP20 | IP20 |
Power switch | 2-poles with inrush limiter |
2-poles with inrush limiter |
2-poles with inrush limiter |
2-poles with inrush limiter |
2-poles with inrush limiter |
Energize mA | <0,3 | <0,3 | <0,5 | <0,5 | <0,5 |
Test voltage kV | 4 | 4 | 4 | 4 | 4 |
PRI-Shield kV | 2 | 2 | 2 | 2 | 2 |
SEC-Shield kV | 2 | 2 | 2 | 2 | 2 |
Ambient temperature | +40°C | +40°C | +40°C | +40°C | +40°C |
Housing | Steel sheet | Steel sheet | Steel sheet | Steel sheet | Steel sheet |
Dimension mm(BxHxT) | 165x110x220 | 165x110x220 | 250x165x310 | 250x165x370 | 250x165x370 |
Mass kg | 6 | 6,6 | 10 | 17 | 26,7 |
Aprobation | CE, RoHS | CE, RoHS | CE, RoHS | CE, RoHS |