Turoidal Isolating Transformers According DIN EN 60601

General Technical Data

Protection degree IP00 
Isolating class B
Ambient temperature max. +45°C
Test voltage shaft - winding 4 kV
Frequenzy 50/60 Hz
Nominal input voltage 230 V
Altitude 1000m
Aprobation CE,RoHs, EAC
Design according DIN EN 60601-2
 
Selection RKT Input Voltage 230 V
Power in VA Sec V
10 V
Sec V
2x6 V
Sec V
2x12 V
Sec V
2x15 V
Sec V
2x18 V
Sec V
2x24 V
Sec V
2x30 V
Sec V
2x48 V
Sec V
2x60 V
Sec V
2x115 V
20 X X X X X          
30 X X X X X          
50   X X X X          
75   X X X X X X      
100     X X X X X      
150     X X X X X      
200     X X X          
250       X X X X      
300     X   X X        
400     X     X   X   X
500             X X X X
630               X X X
800               X X X
1000                 X X
1200                 X X
1600                   X
Dimensions RKT
Power in VA ∅ in mm High in mm Open circuit voltage increase in % Mass in kg
20 61 32 18 0,09
30 74 30 16 0,16
50 79 37 14 0,7
75 89 38 11 1
100 93 42 10 1,25
150 97 51 9 1,75
200 115 53 7 2,3
250 118 55 7 2,5
300 121 63 6 1,23
400 131 64 6 3,5
500 143 65 5 4,3
630 145 77 5 5,1
800 155 72 4 5,5
1000 165 72 3 6,6
1200 195 72 3 7,65
1600 199 98 3 9,8